close



 

 

* Poor contact of cup to cup_connector (the black plastic thing where CE cup sits) on cup stand. 2 metal contact points: the fat one is + pole  and the thin one is - pole , if  negative pole is  bent outward two much, poor contact might occur and when a cup is put on the cup connector, computer cannot detect the cup. 

 

Remedy:  use a pliers to bend the negative pole "inward" just a little bit and make the wire of the cup have good contact to the poles.

 

 

 

* Poor contact of  A/D card to its PCI slot will show

 

1.  room temperature even if there is no cup on the cupstand.  or,

 

2. a jagged, zig-zag  curve instead of a smooth curve.

 

* Poor contact of  Temp signal processing card of its 5V  power suppy connector will disply -80C on the screen.

 

 

 

"Search card error !" will pops up if this A/D card does not have good contact to its slot. In this case, you can do some adjustment it this card. It will get loose after several months, or years of use because of the small vibration from the cooling fan inside the computer.

 

arrow
arrow
    全站熱搜
    創作者介紹
    創作者 吳老師 的頭像
    吳老師

    leometwu的部落格

    吳老師 發表在 痞客邦 留言(0) 人氣()