目前分類:CE_Meter (34)
- Apr 20 Mon 2009 16:44
Zig-Zagging of cooling curve 冷卻曲線不平順
- Jul 22 Tue 2008 20:58
CE_Meter Catalogue
- Jul 05 Sat 2008 09:56
Correcting C% Si% 調整 補正
- Jun 13 Fri 2008 11:47
Sample Curves 曲線範例
- Jun 10 Tue 2008 10:56
Instruction of using CE_Computer
- Jun 10 Tue 2008 10:46
CE Meter and Cupstand 杯座
- Jun 06 Fri 2008 16:59
Possible Causes of malfunction 故障可能原因
- Feb 17 Sun 2008 21:06
冷卻曲線 cooling curve
- Nov 21 Wed 2007 23:33
CE = 等效碳含量
- Nov 01 Thu 2007 10:14
問題曲線 ( D ) faulty curves
Occasional Rugged cooling rate curve is a sign of poor contact of cup. Test results should be discarded because FC number is way too low. Pour another cup right away.
- Oct 26 Fri 2007 09:45
問題冷卻曲線 faulty cooling curves ( III )
Cooling rate curve (green) is rugged (not smooth)----an indication of poor contact of cup to cup connector. This caused the entire cooling curve to drop 20~30 C, resulting in a much lower FC number than it should have been. Test results should be discarded.
- Oct 07 Sun 2007 22:31
冷卻曲線範例 ( II ) Sample curves
A: wrong position of TL. B: Correct position ( Should be應為 FC270 rather than而非 FC301). C: slope of cooling curve before TL is very steep. D: slope of cooling curve after TL is much gentler than C. From the slope of cooling curve, you should be able to tell which stage of solidification a curve segment represents. 從曲線斜率 要能夠判斷凝固進行是到C, D哪個階段 才算師傅. 有時電腦會研判錯誤, 抓錯初晶或共晶點, 此時人為判定需介入. Human judgement should get in when computer fails to determine TL, TE correctly.
- Sep 10 Mon 2007 19:43
冷卻曲線 cooling curve ( 一 )
- Sep 09 Sun 2007 12:48
里歐班LeoMet Class: 冷激試片 Chill test is a must in daily production.